TH
- 160 |
|
 |
- Developed
model of TH140
- On-Board
memory holds 240-1000 groups of
data
- Automatic
identification Impact Device and
test direction (Except G)
- Time
and date setting, auto-clock
- Integral
thermal printer, print all test
results and histogram
- Li
Battery, Low voltage indication
and sound alarm
|
|
|
- Dataview
for PC operation
- Software
data and upper lower limits
setting and sound alarm
- Software
to connect with PC
- Large
LCD with back-light, showing all
functions and parameters
- Direct
Display of tensile strength (U.T.S.)
- For
all metallic materials
- Wide
measuring range (see next page)
- Six
Impact Device are available for
special applications
|
|
Measuring range |
Material |
Hardness
Scale |
D/DC
LD: 170-900 |
D+15
LD+15: 330-900 |
C
LC:
350-960 |
DL
LDL: 560-950 |
G
LG: 200-750 |
Steel
and
Cast steel |
HRC |
20-68.4 |
19.3-67.9 |
20-69.5 |
20.6-68.2 |
|
HRB |
38.4-99.8 |
|
|
37-99.9 |
47.7-99.9 |
HRA |
|
|
|
|
|
HB |
81-654 |
80.638 |
80.683 |
81-6446 |
90-646 |
HV |
81-955 |
80.937 |
80.996 |
80-950 |
|
HS |
32.5-99.5 |
33.3-99.3 |
31.8-102.1 |
30.6-96.8 |
|
Stainless |
HRB |
46.5-101.7 |
|
|
|
|
HB |
85-655 |
|
|
|
|
HV |
85-802 |
|
|
|
|
CWT/st |
HRC |
20.4-67.1 |
19.8-68.2 |
20.7-68.2 |
|
|
HV |
80-898 |
80-935 |
100-941 |
|
|
GC.
Iron |
HRC |
|
|
|
|
|
HB |
93-334 |
|
|
|
92-326 |
HV |
|
|
|
|
|
NC.
Iron |
HRC |
|
|
|
|
|
HB |
131-387 |
|
|
|
127-364 |
HV |
|
|
|
|
|
C.
Alum |
HB |
40-173 |
|
|
|
|
HRB |
13.5-95.3 |
|
|
|
|
Bronze |
HB |
60-290 |
|
|
|
|
Copper |
HB |
45-315 |
|
|
|
|
|
|
Technical
Specifications |
|
Hardness
Scale |
HL,
HRC, HRB, HV, HB, HS |
Memory |
240-1000
groups (Impact times: 31-1) |
Measuring
range |
See
next page |
Tensile
strength U.T.S. range |
374-2652
MPA |
Accuracy |
+
6HLD (760 + 30HLD) error of displayed
value 6HLD (760 + 30HLD)
repeatability of displayed value |
Statistics
function |
Average
/ Max / Min value calculation, limits
setting and alarm |
Standard
Impact Device |
D |
Optional
Impact Device |
DC/DL/C/D+15/G |
Max.
Work piece Hardness |
996HV
(For Impact Devices D/DC/DL/D+15/C) 646HB
(for Impact Device G) |
Min.
Radius of Work piece (Convex/concave) |
R
min = 50mm (with support ring r min = 10mm) |
Min.
Work piece weight |
2-5
kg on stable support 0.05-5 kg with compact
coupling |
Min.
Work piece thickness |
5
mm (Impact Device D/DC/DL/D+15)
1mm (Impact Device C)
10mm (Impact Device G) |
Min.
thickness of hardness Surface |
0.8
mm |
Power |
Rechargeable
Li Battery, pile for TH160 |
Continuous
Working time |
About
100h (no printing and backlight) |
Charging
time |
2.5-4
hours |
Operating
temperature |
0-40°C |
Relative
humidity |
<
90% |
Overall
dimensions |
230
X 90 S 46.5 mm |
Weight |
420g
(Including Impact Device and printer) |
|
|
 |
|
Standard
delivery |
- Main
Unit with removable printer
- Impact
Device
- Test
block with HLD value
- Charger
- Cleaning
brush
- Time
certificate
- Instruction
manual
- Carrying
case
|
|
1
1
1
1
1
1
1 |
|
|
Optional
accessories
|
- Printing
paper
- Special Impact Device
- Support Rings
|
|
|
|
|
|
|
|
|